Free temperature data logger calibration log template
Free temperature data logger calibration log covering ice-point, boiling-point and 3-point comparison per Food Standard 3.2.2 and AS ISO/IEC 17025.
Download free PDF template
Use in MapTrack →Trusted by teams across Australia and New Zealand
We respect your privacy. Unsubscribe anytime.
Updated 26 Oct 2026
Temperature Data Logger Calibration Log Template
Calibration log | Per Food Standard 3.2.2, AS ISO/IEC 17025 and AS 2853
Reference standards & setup
Ice point check
Boiling point check
Working range test points
Drift analysis
Certification & labelling
How to use: download the PDF, print or complete digitally on any device.
- PDF format, ready to print or fill on screen
- Use as-is or customise to suit your operation
- Go digital in MapTrack for photos, reminders and audit trails
Or set it up in MapTrack
- Fill it in on a phone, attach photos to any check
- Automatic reports and audit trails
- Upload the PDF and MapTrack drafts the form for you
Free trial. No credit card. Personal email accepted.
Don't just download it - use it live. Start a free workspace, upload this temperature data logger calibration log template and MapTrack drafts it as a form your crew can run from a phone.
Co-Founder & Commercial Director
Short answer
A temperature data logger calibration log records each logger's identity, the traceable reference probe, and the as-found and as-left readings at each calibration point, so sensor drift is caught. Food, pharmacy and cold chain teams use it, usually every 12 months as a common practice. Food Standard 3.2.2 clause 22 requires premises handling potentially hazardous food to have a thermometer accurate to plus or minus 1 degree C.
Checked against Food Standard 3.2.2 (Compilation F2025C00834, effective 2025-08-13, registered 2025-09-05; current as at 2026-10-26 (Federal Register of Legislation)) on by the MapTrack templates team. How we check templates
Key takeaways
- The MapTrack temperature data logger calibration log template contains 30 checks across 6 sections. Sections covered: Reference standards & setup, Ice point check, Boiling point check, Working range test points, Drift analysis, Certification & labelling.
- Food Standard 3.2.2 clause 22 sets the only legal accuracy figure here: plus or minus 1 degree C for the food premises thermometer.
- Temperature control means 5 degrees C or below, or 60 degrees C or above, unless another temperature is shown to be safe (clause 1).
- A 12-month interval is common practice for food loggers; set pharmaceutical intervals from your quality system and the manufacturer, and shorten either when drift nears tolerance.
- Record as-found readings before adjusting: an out-of-tolerance logger puts every reading since its last calibration in question.
- Use a reference probe with a current traceable certificate; AS ISO/IEC 17025:2018 is the standard calibration laboratories are accredited against.
Enter your email below to download this temperature data logger calibration log template as a ready-to-use PDF.
Used by construction, mining and field service teams
What is a temperature data logger calibration log template?
A temperature data logger calibration log is a formal record used to document the periodic calibration of an NTC, PT100 or thermocouple data logger against traceable reference probes. The log captures the instrument identity, the reference probe used, the data logger sensor type, the rated accuracy class (typically plus or minus 0.5 degrees Celsius for food cold chain duty and plus or minus 0.2 degrees Celsius for pharmaceutical cold chain duty), and the as-found and as-left readings at each calibration point. Standard practice covers an ice-point check at 0 degrees Celsius, a boiling-point check at 100 degrees Celsius corrected for calibrated barometric pressure, and a 3-point comparison sweep against a reference probe across the rated working range. Sensor drift over the 12 month interval is then trended against previous calibration sheets to flag a logger that has aged out of its accuracy class.
From a regulatory angle, temperature data logger calibration sits inside a tight framework that food safety auditors, pharmaceutical quality teams and HACCP coordinators rely on for cold chain integrity evidence. Food Standard 3.2.2 in the Australia New Zealand Food Standards Code requires food businesses that handle potentially hazardous food to have a readily accessible temperature measuring device accurate to plus or minus 1 degree Celsius (clause 22). NATA accreditation provides the practical pathway to a traceable calibration certificate. AS ISO/IEC 17025 sets the international competence standard for the calibration laboratory or in-house procedure that issues the certificate, with traceability through the National Measurement Institute. TGA cold chain guidelines require monitoring equipment to be calibrated and traceable for vaccine, blood bank and pharmacy storage. A documented calibration log is the artefact food safety auditors, TGA inspectors and HACCP coordinators will look for first when they question the integrity of a temperature reading on a food cold chain, pharma cold chain or vaccine distribution run.
Learn more about compliance and inspections in MapTrack.
Get the free template
Enter your email above to download the full temperature data logger calibration log template as a PDF.
Benefits of using this temperature data logger calibration log template
- Cold chain integrity assurance: documented calibration confirms data logger readings are accurate before they are used as cold chain evidence under Food Standard 3.2.2 guidelines
- HACCP compliance: critical control point monitoring requires data loggers to be calibrated and traceable before readings can be used for cooking, cooling or chilled storage
- Regulatory compliance: AS ISO/IEC 17025 and NATA accreditation require monitoring equipment to be calibrated against a traceable reference, with records maintained for audit and investigation
- Detect sensor drift early: trended as-found readings against the reference probe reveal NTC, PT100 or thermocouple drift before the logger falls outside its accuracy class
- Ice-point and boiling-point verification: the dual fixed-point check intercepts sensor offset or span error before it skews a reading at the critical temperature band
- 3-point comparison rigour: comparison against a reference probe at low, mid and high points confirms linearity across the working envelope, not just at fixed points
- Sensor-type discipline: separate calibration records for NTC, PT100 and thermocouple loggers force the right calibration procedure to be applied to the right sensor type
- Due diligence record: after a cold chain breach or vaccine spoilage event, a stamped log proves the data logger was verified at monitoring time
Benefits of digitising forms in MapTrack
When you digitise temperature data logger calibration logs in MapTrack, you get:
- Field users can easily scan a QR code to complete a form on mobile. Unlimited users.
- Email the right people when a form is submitted, with an automation you set up.
- File every completed form against the asset, location or person it covers.
- Download any completed form as a PDF, with failed answers listed at the top.
- Fill in forms offline; they upload when the device is back online.
- Build conditional logic (show or hide questions based on answers).
- Take pictures or attach photos. Not possible with a paper-based form.
- Electronic signatures.
- Edit forms later without reprinting.
- Set who can view, fill in or edit forms with user roles.
- Upload a paper form (PDF, Word, Excel or a photo) and MapTrack's AI builds a digital version for you to check.
- Assign audits and send reminder emails to the people doing them.
- Keep every audit, inspection and sign-off against the asset or site it covers.
- Track corrective actions as work orders from finding to close-out, with each change in the activity log.
Book a demo to see how MapTrack handles temperature data logger calibration logs.
Try MapTrack free
Full access to every feature. No credit card required. Per-asset pricing so you scale as your fleet grows.
- No credit card required
1-2 days/week saved
“Bloody amazing! We used to spend 1-2 days a week tracking and managing our generators alone.”
Steve McAllister
Asset Coordinator, Saunders International
What to include in a temperature data logger calibration log template
This temperature data logger calibration log template covers 11 key areas:
- Instrument details: make, model, serial number, asset/ID, sensor type (NTC, PT100 or thermocouple), accuracy class and rated working range
- Reference probe details: make, model, serial number and NATA-traceable certificate number of the reference probe, four times better than the data logger accuracy class
- Environmental conditions: ambient temperature, relative humidity and barometric pressure recorded against the calibration event for boiling-point correction
- Ice-point check at 0 degrees Celsius: as-found and as-left logger readings against an ice slurry of distilled water and shaved ice, against accuracy class
- Boiling-point check at 100 degrees Celsius: as-found and as-left logger readings against rolling boiling distilled water corrected for barometric pressure on the day of calibration
- 3-point comparison sweep: as-found and as-left logger readings against the reference probe at three test points across the rated working range envelope
- Sensor drift over 12 months: comparison of current as-found readings against the previous calibration sheet to flag a logger that has aged out of class
- Adjustment notes: details of any zero, span or linearity adjustment made during the calibration, with the as-left readings clearly distinguished from the as-found readings
- Application category sign-off: confirmation that the data logger accuracy class is fit for the application, with a note for category changes since the previous calibration
- Technician identity and sign-off: cold chain coordinator or pharmaceutical quality technician name, qualification number, calibration date and signature confirming the logger is fit for service
- Calibration history register: 8-row table capturing date, calibrated by, next due, overall result and NATA certificate number for each event over instrument life
How to use this temperature data logger calibration log template
- 1. Record the instrument details: fill in make, model, serial number, sensor type, accuracy class and rated working range at the top of page 1, cross-reference against the cold chain instrument register to confirm the data logger identity and application category before any reference probe is connected
- 2. Record the reference probe: capture make, model, serial number and NATA-traceable certificate number of the reference probe, confirm the reference accuracy is at least four times better than the data logger accuracy class and that the reference certificate is current
- 3. Stabilise the logger: allow the data logger and reference probe to sit at the calibration room ambient temperature for at least 30 minutes before testing, log the ambient temperature, relative humidity and barometric pressure, and replace the logger battery if condition warrants it
- 4. Run the ice-point check: prepare an ice slurry from shaved distilled water ice and a small volume of distilled water, immerse both the logger probe and the reference probe to the rated immersion depth, allow 5 minutes for stabilisation and record the as-found logger reading against the reference
- 5. Run the boiling-point check: bring a beaker of distilled water to a rolling boil on a controlled heat plate, immerse both probes to the rated immersion depth, allow 5 minutes for stabilisation and record the as-found logger reading against the reference corrected for the calibrated barometric pressure
- 6. Run the 3-point comparison sweep: place both probes in a calibrated dry block or temperature-controlled bath at the low, mid and high points of the rated working range in turn, allow 5 minutes for stabilisation at each point and record the as-found logger reading against the reference
- 7. Compare against previous calibration: pull the previous calibration sheet from the asset record in MapTrack, compare current as-found readings to the previous as-left readings at each test point and calculate sensor drift over the 12 month interval to flag an aged logger
- 8. Adjust if required and re-test: if any test point falls outside the accuracy class, perform the manufacturer-specified adjustment or send the data logger for factory calibration, then repeat the ice-point, boiling-point and 3-point sweep and record the as-left readings
- 9. Close the record: stamp the calibration sheet with the technician name and qualification number, set the next calibration due date based on the application interval, attach a NATA-traceable calibration sticker to the logger body and load the certificate against the asset in MapTrack
In MapTrack, you can automate compliance tracking and audit trails. Each submission is stored as a timestamped PDF against the asset record.
Get the free templateEnter your email above to download the full temperature data logger calibration log template as a PDF.Back to download formHow often should you complete this calibration log?
Temperature data logger calibration intervals are driven by the cold chain application and the accuracy class required. For food cold chain monitoring under Food Standard 3.2.2 at plus or minus 0.5 degrees Celsius accuracy, the typical interval is 12 months from the previous calibration sticker. For pharmaceutical cold chain duty at plus or minus 0.2 degrees Celsius accuracy under TGA cold chain guidelines, the interval tightens to 6 months because vaccine, blood bank and pharmacy storage tolerances are tighter. For HACCP critical control point monitoring on cooking, cooling and hot-holding loops, the interval is typically 12 months but is often supplemented by a monthly user verification check against a calibrated reference thermometer to catch sensor drift between formal calibrations. Outside the scheduled cadence, a temperature data logger must also be calibrated immediately after a suspected impact, water ingress, battery failure, suspected cold chain breach or extended storage period that puts the sensor outside its rated working range.
Try MapTrack free
Full access to every feature. No credit card required. Per-asset pricing so you scale as your fleet grows.
- No credit card required
Frequently asked questions
Applicable regulatory standards
This template is a starting point. Check the current requirements that apply to your equipment, work and location. The template alone does not establish compliance.
| Standard | Edition | Clauses this template relies on | Source |
|---|---|---|---|
| Australia New Zealand Food Standards Code - Standard 3.2.2 - Food Safety Practices and General Requirements (Australia Only)Sets the plus or minus 1 degree C thermometer accuracy, the temperature control limits, and storage of potentially hazardous food under temperature control. | Compilation F2025C00834, effective 2025-08-13, registered 2025-09-05; current as at 2026-10-26 (Federal Register of Legislation) |
| Food Standards Australia New Zealand (Federal Register of Legislation) |
| General requirements for the competence of testing and calibration laboratoriesCompetence standard for calibration laboratories; in Australia NATA accredits laboratories against it. | AS ISO/IEC 17025:2018 | Whole instrument | Standards AustraliaSold by the publisher |
| Australia New Zealand Food Standards Code - Standard 3.2.2A - Food Safety Management ToolsCategory one businesses keep prescribed temperature and other control records for 3 months unless they show compliance another way; clause 22 accuracy is not prescribed. | F2022L01589, made 1 December 2022, registered 8 December 2022, commenced December 2023; no later compilation listed at 2026-10-26 (Federal Register of Legislation) |
| Food Standards Australia New Zealand (Federal Register of Legislation) |
Australia New Zealand Food Standards Code - Standard 3.2.2 - Food Safety Practices and General Requirements (Australia Only)
Edition: Compilation F2025C00834, effective 2025-08-13, registered 2025-09-05; current as at 2026-10-26 (Federal Register of Legislation)
Sets the plus or minus 1 degree C thermometer accuracy, the temperature control limits, and storage of potentially hazardous food under temperature control.
- cl 22: Premises handling potentially hazardous food must have an accessible thermometer accurate to plus or minus 1 degree C.
- cl 1: Defines terms including temperature control: 5 degrees C or below, 60 degrees C or above, or another temperature shown safe.
- cl 6: Food storage: potentially hazardous food must be stored under temperature control, and kept frozen if it is intended to be stored frozen.
Food Standards Australia New Zealand (Federal Register of Legislation)
General requirements for the competence of testing and calibration laboratories
Edition: AS ISO/IEC 17025:2018
Competence standard for calibration laboratories; in Australia NATA accredits laboratories against it.
Standards AustraliaSold by the publisher
Australia New Zealand Food Standards Code - Standard 3.2.2A - Food Safety Management Tools
Edition: F2022L01589, made 1 December 2022, registered 8 December 2022, commenced December 2023; no later compilation listed at 2026-10-26 (Federal Register of Legislation)
Category one businesses keep prescribed temperature and other control records for 3 months unless they show compliance another way; clause 22 accuracy is not prescribed.
- 3.2.2A-12: Category one businesses must keep records substantiating prescribed food safety controls for 3 months, unless they can otherwise demonstrate compliance to an authorised officer.
Food Standards Australia New Zealand (Federal Register of Legislation)
Each source was read at its publisher and the page checked on 26 October 2026. The publisher's current text is the authority.
Temperature Data Logger Calibration Log Template preview

The numbers behind this temperature data logger calibration log template
The pages below gather the figures behind this kind of check. Each figure on them carries its source and date, so you can quote it in a toolbox talk, a budget case or a board paper without chasing the original report.
Get the free template
Enter your email above to download the full temperature data logger calibration log template as a PDF.
Embed this free template on your website
Run an industry blog, trade association site, or training resource? Drop a preview of this free temperature data logger calibration log template straight into your page. The snippet is self-contained, needs no scripts, and links readers back to the full free template.
<div style="max-width:480px;font-family:system-ui,-apple-system,'Segoe UI',Roboto,sans-serif;border:1px solid #E5E7EB;border-radius:12px;padding:20px;background:#ffffff;">
<p style="font-size:12px;font-weight:700;letter-spacing:0.05em;text-transform:uppercase;color:#071D49;margin:0;">Free template</p>
<p style="font-size:18px;font-weight:700;color:#071D49;margin:6px 0 0;">Temperature Data Logger Calibration Log Template</p>
<ul style="margin:12px 0 0;padding-left:18px;color:#374151;font-size:14px;line-height:1.6;">
<li style="margin:4px 0;">Instrument details: make, model, serial number, asset/ID, sensor type (NTC, PT100 or thermocouple), accuracy class and rated working range</li>
<li style="margin:4px 0;">Reference probe details: make, model, serial number and NATA-traceable certificate number of the reference probe, four times better than the data logger accuracy class</li>
<li style="margin:4px 0;">Environmental conditions: ambient temperature, relative humidity and barometric pressure recorded against the calibration event for boiling-point correction</li>
<li style="margin:4px 0;">Ice-point check at 0 degrees Celsius: as-found and as-left logger readings against an ice slurry of distilled water and shaved ice, against accuracy class</li>
<li style="margin:4px 0;">Boiling-point check at 100 degrees Celsius: as-found and as-left logger readings against rolling boiling distilled water corrected for barometric pressure on the day of calibration</li>
<li style="margin:4px 0;">3-point comparison sweep: as-found and as-left logger readings against the reference probe at three test points across the rated working range envelope</li>
</ul>
<p style="font-size:13px;color:#6B7280;margin:14px 0 0;padding-top:12px;border-top:1px solid #E5E7EB;">Free <a href="https://www.maptrack.com/templates/temperature-data-logger-calibration-log" style="color:#071D49;font-weight:600;text-decoration:none;">Temperature Data Logger Calibration Log Template</a> by MapTrack</p>
</div>Free to reuse, including commercially. This template is published under a Creative Commons Attribution 4.0 licence. Copy it, adapt it, put it on your own site or in your own docs. All we ask is that you credit MapTrack with a link back to this page. The snippet above already includes that credit, so if you use it as-is you’re covered.
Cite this template
Writing about this topic? You’re welcome to quote this checklist. Here’s the wording to use so your readers can find the original.
The MapTrack temperature data logger calibration log template contains 30 checks across 6 sections.
- Short attribution
- MapTrack Temperature Data Logger Calibration Log Template, https://www.maptrack.com/templates/temperature-data-logger-calibration-log
- Reference list
- MapTrack. (2026). Temperature Data Logger Calibration Log Template [Checklist template]. Retrieved from https://www.maptrack.com/templates/temperature-data-logger-calibration-log
Free to reuse with credit under a Creative Commons Attribution 4.0 licence. If you’d rather link straight to it, the page is https://www.maptrack.com/templates/temperature-data-logger-calibration-log.
Need to automate compliance tracking and audit trails?
Register every temperature data logger in MapTrack, attach digital forms, and get a complete history of every inspection, service and compliance record.
Compliance and inspections · All templates · Pricing · Book a demo


